RTD simulator is aimed for calibrating of temperature controllers. It is suitable for calibration laboratories and service centers, where it can be used also as computer controlled resistance decade.
The instrument is supplied from accumulator or via power line adapter. The Control is possible manually or remotely via serial interface RS232 (IEEE488). RTD simulator can be ordered as 19” module for easy assembling into a 19” rack. Module height is 3HE.
Download M-612 Data sheet
Download M-612 Operation manual
Download M-6x2 Brochure
| Resistance range | 16.0000 Ohm - 10.000 kOhm |
| Range of Pt sensor temperature simulation | -200°C - +850°C |
| Range of Ni sensor temperature simulation | -60°C - +300°C |
| Types of sensors | Pt100, Pt200, Pt500, Pt1000, Ni100, Ni1000 |
| Temperature scales | IPTS68, ITS90 |
| Pt sensor standards | DIN (1,385), US (1,392) |
| Ni sensor standards | DIN 43760 (6180) |
| Connection | 2, 3, 4-wire |
| Remote control | RS-232, optionally IEEE488 |
| Power supply | internal ACCU/power line adapter (100-240V/50-60Hz) |
| Dimmensions | W 325 mm, H 111 mm, D 316 mm |
| Temperature | Pt100 | Pt200 | Pt500 | Pt1000 |
|---|---|---|---|---|
| -200 ... 200°C | 0.02°C | 0.02°C | 0.02°C | 0.03°C |
| 200 ... 500°C | 0.03°C | 0.04°C | 0.06°C | 0.15°C |
| 500 ... 850°C | 0.04°C | 0.06°C | 0.15°C | 0.20°C |
| Temperature | Ni100 | Ni1000 |
|---|---|---|
| -60 ... 300°C | 0.02°C | 0.1°C |
| Resistance range | Accuracy |
|---|---|
| 16 - 400 Ohm | 0.003 % + 3 mOhm |
| 400 - 2000 Ohm | 0.005 % |
| 2000 - 10000 Ohm | 0.015 % |
| Power line adapter | 1 pc |
| Cable RS232 (for RS232 version only) | 1 pc |
| Operation manual | 1 pc |
| Application software | CD ROM |
| Cable IEEE488 | IEEE488/IEEE488 2m |
| Spare accumulator 12V 1,9Ah | B-WP1,9 - 12 |
| Bus | M612-V1xxx ... RS232 |
| M612-V2xxx ... IEEE488 | |
| Additional functions | M612-Vx0xx ... none |
| M612-Vx1xx ... with Short / Open | |
| Housing | M612-Vxx0x ... table version |
| M612-Vxx1x ... module 19", 3HE |