The resistance standard set is designed especially for calibration of LCR meters in frequency range from DC to 100 MHz. The standards are equipped with four coaxial BNC connectors. Wide frequency range is achieved by applied HF design using strip line technology. Four-pair terminal connection with 22 mm distance between them enables direct connection to the UUTs without need to apply test leads.
The standards are available with nominal value 10, 50, 100, 500, 1000 Ω in 1% tolerance. Reference positions OPEN and SHORT for easy “zero” compensation are available as well.
Model |
Nominal value |
Max. deviation to nominal value |
Maximal frequency for 0.1% deviation |
Maximal frequency for 1% deviation |
Time constant |
Temperature coefficient |
Maximal test current |
|
|
W |
% |
MHz |
MHz |
s |
ppm/°C |
mA |
|
M530-10R |
10 |
1 |
2 |
200 |
< 2.10-10 |
|< 1 |
40 |
|
M530-50R |
50 |
1 |
10 |
> 500 |
< 2.10-10 |
< 1 |
20 |
|
M530-100R |
100 |
1 |
2 |
200 |
< 5.10-11 |
< 1 |
12 |
|
M530-500R |
500 |
1 |
0.5 |
50 |
< 2.10-10 |
< 25 |
6 |
|
M530-1000R |
1000 |
1 |
0.1 |
10 |
< 2.10-10 |
< 25 |
4 |
|
OPEN |
- |
- |
2 |
200 |
- |
- |
- |
|
SHORT |
0 |
- |
2 |
200 |
- |
- |
100 |